Manufacturer > Abbott > Devices > AMPLATZER™ VASCULAR PLUG (AVP)

AMPLATZER™ VASCULAR PLUG (AVP)

Device-Type

Vascular Plug

Targated Speciality

Cardiovascular

Manufacturer

Abbott


  • Self-expanding device made from a Nitinol wire mesh
  • The device is shipped attached to a 135cm nitinol delivery wire in a hoop dispenser
  • The device is preloaded in a loader
  • No plastic vise is provided
  • Sizes ranging from 4mm – 16mm in 2mm increments


Materials:

  • Plug: Nitinol
  • Delivery wire: Nitinol
  • Marker bands: Platinum
  • Micro screw: Stainless steel


Features and Benefits

Designed to provide precise positioning in short landing zones

  • Single lobe, single layer device
  • Platinum marker bands on both ends
  • Select a device that is ~30-50% larger in diameter than the target vessel diameter

amplatzer vascular plug

Related Device SKUs

* Select 2 or more (upto 4) SKUs to compare.
SKU Name Device SKUs Model VASCULAR PLUG DIAMETER (A) UNCONSTRAINED LENGTH (B) TREATABLE VESSEL DIAMETER RANGE
AVP - 10 mm x 7 mm
9-PLUG-010 10 mm 7 mm 6.5 mm-7.5 mm
AVP - 12 mm x 8 mm
9-PLUG-012 12 mm 8 mm 8.0 mm-9.0 mm
AVP - 14 mm x 8 mm
9-PLUG-014 14 mm 8 mm 9.5 mm-11.0 mm
AVP - 16 mm x 8 mm
9-PLUG-016 16 mm 8 mm 10.5 mm-12.5 mm
AVP - 4 mm x 7 mm
9-PLUG-004 4 mm 7 mm 2.5 mm-3.0 mm
AVP - 6 mm x 7 mm
9-PLUG-006 6 mm 7 mm 4.0 mm-4.5 mm
AVP - 8 mm x 7 mm
9-PLUG-008 8 mm 7 mm 5.5 mm-6.0 mm
device-thumbnail
AVP - 10 mm x 7 mm
  • Model : 9-PLUG-010
  • Procedure : -
  • VASCULAR PLUG DIAMETER (A) : 10 mm
  • UNCONSTRAINED LENGTH (B) : 7 mm
  • TREATABLE VESSEL DIAMETER RANGE : 6.5 mm-7.5 mm
device-thumbnail
AVP - 12 mm x 8 mm
  • Model : 9-PLUG-012
  • Procedure : -
  • VASCULAR PLUG DIAMETER (A) : 12 mm
  • UNCONSTRAINED LENGTH (B) : 8 mm
  • TREATABLE VESSEL DIAMETER RANGE : 8.0 mm-9.0 mm
device-thumbnail
AVP - 14 mm x 8 mm
  • Model : 9-PLUG-014
  • Procedure : -
  • VASCULAR PLUG DIAMETER (A) : 14 mm
  • UNCONSTRAINED LENGTH (B) : 8 mm
  • TREATABLE VESSEL DIAMETER RANGE : 9.5 mm-11.0 mm
device-thumbnail
AVP - 16 mm x 8 mm
  • Model : 9-PLUG-016
  • Procedure : -
  • VASCULAR PLUG DIAMETER (A) : 16 mm
  • UNCONSTRAINED LENGTH (B) : 8 mm
  • TREATABLE VESSEL DIAMETER RANGE : 10.5 mm-12.5 mm
device-thumbnail
AVP - 4 mm x 7 mm
  • Model : 9-PLUG-004
  • Procedure : -
  • VASCULAR PLUG DIAMETER (A) : 4 mm
  • UNCONSTRAINED LENGTH (B) : 7 mm
  • TREATABLE VESSEL DIAMETER RANGE : 2.5 mm-3.0 mm
device-thumbnail
AVP - 6 mm x 7 mm
  • Model : 9-PLUG-006
  • Procedure : -
  • VASCULAR PLUG DIAMETER (A) : 6 mm
  • UNCONSTRAINED LENGTH (B) : 7 mm
  • TREATABLE VESSEL DIAMETER RANGE : 4.0 mm-4.5 mm
device-thumbnail
AVP - 8 mm x 7 mm
  • Model : 9-PLUG-008
  • Procedure : -
  • VASCULAR PLUG DIAMETER (A) : 8 mm
  • UNCONSTRAINED LENGTH (B) : 7 mm
  • TREATABLE VESSEL DIAMETER RANGE : 5.5 mm-6.0 mm