EmeryGlide™ MR Conditional Guidewire

Device-Type

Mandril/Specialty Guidewires

Targated Speciality

Endovascular


Safe by Design

  • The EmeryGlide is constructed from a high strength core composite of glass fibers and polymers, protected by a double winding of aramid fibers embedded in polymer.
  • There is a full body (tip to end) PTFE sleeve that minimizes intravascular friction and protects the tip and shaft from breakage.
  • The distal tip of the EmeryGlide is marked with 3 discrete markers containing different nanoparticles that enable visibility under MRI as well as fluoroscopy.
  • The passive markers are positioned 0cm, 2cm and 4cm from the straight or angled tip.

Features and Benefits

MR Conditional Statement

Non‐clinical testing has demonstrated that the EmeryGlide MR Guidewire is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:

  • Static magnetic field of 1.5 or 3.0 T
  • Maximum spatial field gradient of 3600 Gauss/cm (36.0 T/m) for 1.5 T systems
  • Maximum spatial field gradient of 1800 Gauss/cm (18.0 T/m) for 3.0 T systems
  • Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 4.0 W/kg (First Level Operating Mode at 1.5 T)
  • Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 4.0 W/kg (First Level Operating Mode at 3.0 T)

Safety informations

Instructions For Use

Carefully read the following warnings, precautions and directions prior to use. Failure to do so may result in improper use of this device which could cause the following complications:

- Shearing of the MR Guidewire.

- Release of fragments of the MR Guidewire.

- Vessel trauma.

Contraindication

The EMERYGLIDE™ MR Guidewire is not intended for coronary or cerebral vasculature.  



Potential adverse events

Complications that may be related to the use of EMERYGLIDE™ MR Guidewire can include but may not be limited to the following:

  • Thrombus
  • Emboli
  • Arterial or venous wall damage
  • Plaque dislodgement
  • Hematoma at the puncture site
  • Infection
  • Perforation
  • Vessel spasm
  • Hemorrhage
  • Vascular thrombosis
  • Allergic reaction
  • Arrhythmia’s

Device Documents

Related Device SKUs

* Select 2 or more (upto 4) SKUs to compare.
SKU Name Device SKUs Model Tip Shape Diameter (in) Length (cm)
e EmeryGlide
CV-G-18035S Straight 0.035 180
e EmeryGlide
CV-G-18035A Angled 0.035 180
device-thumbnail
e EmeryGlide
  • Model : CV-G-18035S
  • Procedure : -
  • Tip Shape : Straight
  • Diameter (in) : 0.035
  • Length (cm) : 180
device-thumbnail
e EmeryGlide
  • Model : CV-G-18035A
  • Procedure : -
  • Tip Shape : Angled
  • Diameter (in) : 0.035
  • Length (cm) : 180