Manufacturer > Merit Medical > Devices > Merit SureCross™ Support Catheters

Merit SureCross™ Support Catheters

Device-Type

Specialty Catheters

Targated Speciality

Endovascular

Manufacturer

Merit Medical


Crossing tight, difficult lesions is often challenging and time consuming. Having a support catheter that offers trackability, pushability and visibility gives you the confidence needed to cross occlusions, even in your most challenging case.


The Merit SureCross Support Catheter is an over-the-wire (OTW) single-lumen catheter with an atraumatic tapered tip. The catheter system is offered in twelve (12) size models that are 0.014”, 0.018”, and 0.035” guidewire compatible with working lengths of 65 cm, 90 cm, 135 cm, and 150 cm. The distal catheter shaft has three (3) radiopaque markers that aid in positioning the catheter and estimating distances. The distal 40 cm outer portion of the catheter has a hydrophilic coating. The proximal end of the catheter has a hub with a female luer connection. The lumen is used to pass the catheter over the appropriate guidewire (as listed in Table 1) or for infusion. The catheter guidewire size and length are printed on the strain relief. 

Features and Benefits

Trackability, Pushability, Visibility

Crossing tight, difficult lesions is often challenging and time consuming. Having a support catheter that offers trackability, pushability and visibility gives you the confidence needed to cross occlusions, even in your most challenging case.


Product Features

Translucent Shaft for visualization of blood to verify placement in vessel lumen

3 Platinum Marker Bands to confirm catheter position and lesion sizing

  • First marker band 2 mm from tip
  • 0.014” and 0.018” second and third marker bands are 15 mm apart
  • 0.035” second and third marker bands are 50 mm apart

40 cm Distal Hydrophilic Coating provides smooth tracking through tortuous anatomy

Low Profile Tapered Tip facilitates crossing difficult lesions

Maximum Infusion Pressure 300psi

Guide Wire Compatibility 0.014”, 0.018”, 0.035”

Multiple Lengths 65 cm, 90 cm, 135 cm, 150 cm

surecross-shaft


SureCross Shaft Technology

Engineered to optimize pushability and flexibility which maximizes guide wire support, trackability and lesion crossing.

surecross-tip-design


SureCross Tip Design

Short tip distance to marker band (2 mm) maximizes accuracy of tip location and enhances tip radiopacity

Internal platinum distal marker band provides a seamless transition zone

Marker band improves critical tip visualization

Robust tapered tip improves crossing and durability

Excellent transition from catheter tip to guide wire

surecross-tip-guidewire-comparison


Tip to Guide Wire Comparison

The low profile, seamless transition from catheter to guide wire provides exceptional crossing support.

Safety informations

INDICATIONS FOR USE

The Merit SureCross Support Catheter is intended for use during coronary and peripheral interventional procedures to guide and support guidewires, traverse discrete portions of the vasculature, allow for guidewire exchanges and provide a conduit for infusion of saline solution, diagnostic contrast agents and therapeutic agents.


WARNINGS

• The Support Catheter is supplied STERILE and for single use only. Do not reprocess or re-sterilize. Reprocessing and re-sterilizing could increase the risk of patient infection and/or compromise device performance.

• The catheter should only be manipulated while observing under fluoroscopy.

• If resistance is encountered at any time during the insertion procedure, do not force passage or torque the catheter. Resistance may cause damage to device or vessel. Carefully withdraw the catheter.

• The catheter should only be advanced or withdrawn over a guidewire.

• Do not exceed the maximum recommended infusion pressure of 300psi.

• This device has not been evaluated for use in the neurovasculature.


PRECAUTIONS

• Preparations should be made and a trained vascular surgical team available in the event conversion to open surgery is required.

• Carefully inspect the package and catheter prior to use to verify no damage occurred during shipment. Do not use if the package or catheter is damaged since the sterility or integrity of the device may be compromised and thus increasing the risk of patient infection and device malfunction.

• To avoid damage to hydrophilic coating, do not wipe the catheter surface with dry gauze.

• To avoid kinking and damaging the catheter, advance slowly in small increments until the proximal end of the guidewire emerges from the catheter.

• To avoid damage to the catheter or vessel, do not advance or withdraw the device without a guidewire in place.

• Do not exceed the maximum recommended infusion pressure of 300 PSI. Higher pressures may cause damage to the catheter and vessels. Refer to Table 2 for flow rates of the catheter.

• Catheter should not be advanced into a vessel having a diameter smaller than the catheter outer diameter. Damage to the vessel or device may occur.

• After use, this product may be a potential biohazard. Handle and dispose of in accordance with acceptable medical practice and applicable local, state, and federal laws and regulations.

• Use the catheter prior to the “Use By” date specified on the package. 


PACKAGING, STERILIZATION AND STORAGE

The device has been sterilized using ethylene oxide (EO) and is supplied sterile and non-pyrogenic. The package label indicates the Use By date. Do not use the device after the Use By date. The device sterility and integrity may be compromised and possibly result in patient infection and/or device malfunction. The device should be stored in a cool, dry and dark place. Do not store catheters where they are directly exposed to organic solvents, ionizing radiation, ultraviolet light and high temperatures (greater than 60°C or 140°F). Rotate inventory so that device is used prior to the Use By date on the package label.


RECOMMENDED ITEMS Each Support Catheter package includes the following:

• Single-use sterile disposable catheter contained in a coiled storage tube. Materials required but not provided are:

• 10 ml syringe (filled with sterile heparinized saline)

• One-Way or Three-Way stopcock

• Appropriately sized exchange guidewire

• Appropriately sized hemostatic introducer sheath or guiding catheter

• Additional Support Catheters

• Sterile, heparinized saline (for flushing catheter)


DEVICE PREPARATION

8.1 Carefully inspect the package and catheter prior to use to verify no damage occurred during shipment. Do not use catheter if either the catheter or packaging is damaged or compromised.

8.2 If the packaging is free of damage, carefully open the pouch and introduce the sterile hoop (with catheter) to the sterile field using sterile techniques.

8.3 Carefully disengage the catheter hub from the hoop connector and withdraw the catheter from the hoop. Carefully inspect the catheter for any signs of damage. If damaged, please discard and use another catheter.

8.4 The device is coated with a lubricious hydrophilic coating. Prior to inserting the catheter, activate the coating by immersing the catheter in normal saline for approximately 30 seconds, or wiping down the catheter shaft with a saturated gauze or sponge. Do not wipe the catheter surface with dry gauze.

8.5 Prepare the catheter by removing the air in the catheter lumen. Connect a 10 ml syringe filled with heparinized saline to the hub luer and flush out the lumen. While flushing the catheter, visually inspect the catheter for any signs of damage or leakage. If any damage or leakage is observed, discard and use a new device.



DEVICE USAGE

9.1 Using standard techniques, place the introducer sheath/guiding catheter and guidewire.

9.2 Place the prepared Support Catheter over the previously inserted guidewire by threading the end of the guidewire through the tip of the catheter. Advance the catheter over the guidewire. Refer to Table 1 for guidance regarding appropriate catheter selection along with introducer sheath, guiding catheter and guidewire sizing.

9.3 The catheter can be introduced percutaneously through the introducer sheath or guiding catheter. Do not attempt to pass the device through a smaller sized sheath introducer than indicated.

9.4 Using fluoroscopic imaging, carefully advance the catheter to the desired location in the vasculature. If resistance is encountered at any time during the insertion procedure, do not force passage or torque the device. The markerbands on the distal end of the catheter will provide visual guidance.

9.5 Using fluoroscopic imaging manipulate the guidewire and catheter in order to achieve the desired position. Do not torque the catheter. Also, do not advance or withdraw the catheter without the guidewire in place and extending beyond the distal end of the catheter.

9.6 When removing the catheter, use fluoroscopic imaging and make sure to maintain guidewire position. 

Potential adverse events

Complications may occur with the use of any support catheter or during any catheterization procedure. Therefore, only physicians trained in vascular surgery, interventional radiology or cardiology, and who have completed training or have experience with support catheters and associated devices should consider using this device. Possible complications associated with this type of procedure include but are not limited to the following:

• Access site complications

• Acute or total occlusion of the vessel

• Arterial dissection

• Arterial spasm

• Arterial thrombosis

• Catheter fracture with tip separation and distal embolization

• Death

• Distal embolization (air, blood clots or plaque)

• Hemorrhage or hematoma

• Hypo/hypertension

• Intimal disruption

• Local or systemic infection including sepsis

• Myocardial Infarction

• Perforation and vessel rupture

• Surgical intervention

• Thrombus formation/thrombosis

Device Documents

Related Device SKUs

* Select 2 or more (upto 4) SKUs to compare.
SKU Name Device SKUs Model Color Guide Wire Compatibility Minimum Guiding Catheter(F) Working Length(cm) 2nd and 3rd Marker Band Spacing(MM) Hydrophilic Coating Length(cm)
SureCross Support Catheter
SC1465   0.014" 4 65 15 40
SureCross Support Catheter
SC1490   0.014" 4 90 15 40
SureCross Support Catheter
SC14135   0.014" 4 135 15 40
SureCross Support Catheter
SC14150   0.014" 4 150 15 40
SureCross Support Catheter
SC1865   0.018” 4 65 15 40
SureCross Support Catheter
SC1890   0.018” 4 90 15 40
SureCross Support Catheter
SC18135   0.018” 4 135 15 40
SureCross Support Catheter
SC18150   0.018” 4 150 15 40
SureCross Support Catheter
SC3565   0.035” 5 65 50 40
SureCross Support Catheter
SC3590   0.035” 5 90 50 40
SureCross Support Catheter
SC35150   0.035” 5 150 50 40
  • Model : SC1465
  • Procedure : -
  • Guide Wire Compatibility : 0.014"
  • Minimum Guiding Catheter(F) : 5
  • Minimum Guiding Catheter(F) : 4
  • Working Length(cm) : 65
  • 2nd and 3rd Marker Band Spacing(MM) : 15
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC1490
  • Procedure : -
  • Guide Wire Compatibility : 0.014"
  • Minimum Guiding Catheter(F) : 5
  • Minimum Guiding Catheter(F) : 4
  • Working Length(cm) : 90
  • 2nd and 3rd Marker Band Spacing(MM) : 15
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC14135
  • Procedure : -
  • Guide Wire Compatibility : 0.014"
  • Minimum Guiding Catheter(F) : 5
  • Minimum Guiding Catheter(F) : 4
  • Working Length(cm) : 135
  • 2nd and 3rd Marker Band Spacing(MM) : 15
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC14150
  • Procedure : -
  • Guide Wire Compatibility : 0.014"
  • Minimum Guiding Catheter(F) : 5
  • Minimum Guiding Catheter(F) : 4
  • Working Length(cm) : 150
  • 2nd and 3rd Marker Band Spacing(MM) : 15
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC1865
  • Procedure : -
  • Guide Wire Compatibility : 0.018”
  • Minimum Guiding Catheter(F) : 5
  • Minimum Guiding Catheter(F) : 4
  • Working Length(cm) : 65
  • 2nd and 3rd Marker Band Spacing(MM) : 15
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC1890
  • Procedure : -
  • Guide Wire Compatibility : 0.018”
  • Minimum Guiding Catheter(F) : 5
  • Minimum Guiding Catheter(F) : 4
  • Working Length(cm) : 90
  • 2nd and 3rd Marker Band Spacing(MM) : 15
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC18135
  • Procedure : -
  • Guide Wire Compatibility : 0.018”
  • Minimum Guiding Catheter(F) : 5
  • Minimum Guiding Catheter(F) : 4
  • Working Length(cm) : 135
  • 2nd and 3rd Marker Band Spacing(MM) : 15
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC18150
  • Procedure : -
  • Guide Wire Compatibility : 0.018”
  • Minimum Guiding Catheter(F) : 5
  • Minimum Guiding Catheter(F) : 4
  • Working Length(cm) : 150
  • 2nd and 3rd Marker Band Spacing(MM) : 15
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC3565
  • Procedure : -
  • Guide Wire Compatibility : 0.035”
  • Minimum Guiding Catheter(F) : 6
  • Minimum Guiding Catheter(F) : 5
  • Working Length(cm) : 65
  • 2nd and 3rd Marker Band Spacing(MM) : 50
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC3590
  • Procedure : -
  • Guide Wire Compatibility : 0.035”
  • Minimum Guiding Catheter(F) : 6
  • Minimum Guiding Catheter(F) : 5
  • Working Length(cm) : 90
  • 2nd and 3rd Marker Band Spacing(MM) : 50
  • Hydrophilic Coating Length(cm) : 40
  • Model : SC35150
  • Procedure : -
  • Guide Wire Compatibility : 0.035”
  • Minimum Guiding Catheter(F) : 6
  • Minimum Guiding Catheter(F) : 5
  • Working Length(cm) : 150
  • 2nd and 3rd Marker Band Spacing(MM) : 50
  • Hydrophilic Coating Length(cm) : 40

Questions & Answers

Don't see the answer you're looking for ?