Manufacturer > Abbott > Devices > PressureWire™ X Guidewire

PressureWire™ X Guidewire

Device-Type

FFR Technologies

Manufacturer

Abbott


Abbott’s newest generation of the PressureWire™ X Guidewire offers objective decisions with a simplified workflow.

The innovative PressureWire™ X Guidewire—the world’s only wireless physiology wire2—can measure pressure and temperature to calculate Abbott's Resting Full-Cycle Ratio (RFR), Fractional Flow Reserve (FFR), Index of Microvascular Resistance (IMR), and Coronary Flow Reserve (CFR). The guidewire’s fully integrated, secure, wireless measurements are integral to a cardiac cath lab’s clinical physiology routine.


The PressureWire™ X Guidewire In The Cath Lab

PressureWire™ X setup specifications



DRIFT


BE CONFIDENT IN YOUR RESTING INDEX DECISION-MAKING with Abbott’s Pressure Wire™ X Guidewire: Abbott’s innovative sensor housing designs offers <1 mmHg/30 minutes with an 84% reduction in drift due to microbubbles.

Drift is an issue that can lead to misclassification of stenosis.

There are two types of drift: wire-related (true drift) and procedure-related (apparent drift). While there are multiple causes of procedure-related drift, microbubbles (from residual air in device) are a major cause of drift. Clinically significant drift is classified as ≥3 mmHg

Drift affects the accuracy of most pressure measurement devices. Accuracy is particularly critical for resting indices, as clinical decisions from resting indices are more impacted by drift than FFR. Compared to the hyperemic state, smaller separation between Pa and Pd at rest means it is critical to detect even the slightest difference of pressure.


Features and Benefits

Objective Decisions

  • Innovative sensor housing design offers <1 mmHg/30 min with reduction in drift due to microbubbles5
  • Objective decision-making with both RFR and FFR options
  • RFR has been studied in over 3,500 lesions and 2,000 patients6-10 to show diagnostic equivalence to iFR‡.

84% reduced drift with the latest PressureWire X Guidewire compared to its previous sensor housing design


Simplified Workflow

  • The PressureWire™ X Guidewire is the world’s only wireless physiology wire2
  • Wireless: from zero to go in 2 easy steps2
  • Ideal for the modern radial approach­
  • Compatible with the QUANTIEN™ Measurement System with an easy touchscreen workflow
  • Compatible with OPTIS™ Imaging Systems for imaging and physiology needs on one platform
  • Compatible with the Coroventis‡ CoroFlow‡ Cardiovascular System for assessment of both epicardial disease and microvascular dysfunction11

Wireless PressureWire X Guidewire

2 Steps to Connect the Wireless
PressureWire™ X Guidewire1
  1. Zero. Zero the PressureWire™ X Guidewire by turning on the transmitter.
  2. Go. The transmitter’s function indicator shall display a stable green light when it is successfully zeroed.

Recommended Compatible Devices

The PressureWire™ X Guidewire is compatible with the following:


Safety informations

Indications: 

The PressureWire™ X Guidewire is indicated to direct a catheter through a blood vessel and to measure physiological parameters in the heart and in the coronary and peripheral blood vessels. Physiological parameters include blood pressure. The PressureWire™ X Guidewire can also measure blood temperature.


Contraindications: This guidewire is contraindicated for use in the cerebral vasculature.


Warnings:

  • No modification of this device is allowed.
  • The PressureWire™ X Guidewire is supplied sterile. Discard the guidewire if the pouch is opened or damaged, compromising the sterile barrier. The guidewire is designed for single use only and shall not be reused or resterilized. Adverse effects of using a non-sterile or resterilized guidewire may include, but are not limited to:
  • Local and/or systemic infection
  • Mechanical damage
  • Inaccurate readings
  • Observe all guidewire movements. Whenever the guidewire is moved or torqued, the tip movement should be examined under fluoroscopy. Never push, withdraw, or torque the guidewire if it meets resistance or without observing corresponding movement of the tip, otherwise vessel/ventricle trauma may occur.
  • Torquing or excessive manipulation of the guidewire in a sharp bend, against resistance, or repeated attempts to cross a total vessel occlusion may:
  • Cause dissection or perforation of blood vessels
  • Cause vessel spasm
  • Damage and/or fracture the guidewire
  • When introducing the guidewire, flush the catheter and administer anticoagulation as for a standard catheterization procedure or clotting may occur.
  • Do not use the guidewire in the ventricles if the patient has a prosthetic mechanical or biological valve. It may result in damage to both the prosthesis and the guidewire, which may cause injury or death.
  • Use of the PressureWire™ X Guidewire in conjunction with interventional devices with a short rapid exchange may result in a folded or fractured guidewire.
  • High frequency surgical devices must not be used on a patient at the same time as the guidewire.


Precautions:

  • The PressureWire™ X Guidewire is a delicate instrument and should be handled carefully.
  • Make sure that the transmitter is kept dry to ensure accurate pressure and/or temperature readings. Inaccurate readings may necessitate device replacement.
  • Do not use the guidewire in conjunction with atherectomy catheters. It may damage the guidewire.
  • Do not withdraw or manipulate the guidewire in a sharp-edged object. It may result in abrasion of the guidewire coating.
  • Factors that may affect the accuracy of the diagnostic information include, but are not limited to:
  • Improper placement of the aortic pressure sensor.
  • Failure to achieve maximum coronary and myocardial hyperemia.
  • Blood flow affected by the position of interventional devices, such as balloon catheters.
  • Guidewire readings may be affected by defibrillation. Rezero the guidewire after defibrillation use.
  • Do not measure pressure when the guidewire sensor element is in a sharp bend or in contact with atrial or ventricular walls. It might result in pressure artifacts.
  • Do not use the PressureWire™ X Guidewire together with another guidewire, for so called jailed wire technique, due to difficulty in guidewire withdrawal and possible guidewire entrapment.
  • Store at room temperature in a dry and dark place.

Potential adverse events

Potential complications which may be encountered during all catheterization procedures include, but are not limited to: vessel dissection or occlusion, perforation, embolus, spasm, local and/or systemic infection, pneumothorax, congestive heart failure, myocardial infarction, hypotension, chest pain, renal insufficiency, serious arrhythmias, or death.

Related Device SKUs

* Select 2 or more (upto 4) SKUs to compare.

Questions & Answers

Don't see the answer you're looking for ?