PTS-X Sizing Balloon

Device-Type

Specialty Balloons

Targated Speciality

Cardiovascular


Designed for use in patients with cardiovascular defects where accurate measurements of the defect are important to select the appropriately sized occluder device. The PTS-X™ brand provides the additional usable length and benefit of a braided inner lumen for optimal pushability, trackability and kink-resistance.

Features and Benefits

Short balloon tapers for ease of deliverability and withdrawal into the introducer.

Comprehensive size offering with diameters ranging from 20mm – 40mm.

Four radiopaque marker bands to enable accurate defect sizing.

Braided inner lumen for optimal pushability, trackability and kink-resistance.

Recommended Compatible Devices

The PTS-X™ is compatible with the D’VILL™ Introducer (65cm).

Safety informations

WARNINGS:

  • CAUTION: Do not exceed the RBP. An inflation device with pressure gauge is recommended to monitor pressure. Pressure in excess of the RBP can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.
  • Use only appropriate balloon inflation medium. Do not use air or gaseous medium to inflate the balloon.
  • Do not advance the guidewire, balloon dilatation catheter, or any other component if resistance is met, without first determining the cause and taking remedial action.
  • This catheter is not recommended for pressure measurement or fluid injection.
  • Do not remove the guidewire from the catheter at any time during the procedure.
  • This device is intended for single use only. Do not resterilize and/or reuse it, as this can potentially result in compromised device performance and increased risk of cross contamination.


PRECAUTIONS:

• One should always select a PTS-X diameter larger than the unstretched defect diameter, i.e., TEE ASD size 12mm - select 20 or 25 mm PTS-X.

• Caution should be used when inflating the balloon, over inflation can cause trauma and overstretching of the septum.

• Sizing procedures should be conducted under fluoroscopic guidance with appropriate x-ray equipment.

• Guidewires are delicate instruments. Care should be exercised while handling to help prevent the possibility of breakage.

• Careful attention must be paid to the maintenance of tight catheter connections and aspiration before proceeding to avoid air introduction into the system.

• Under no circumstances should any portion of the catheter system be advanced against resistance. The cause of the resistance should be identified with fluoroscopy and action taken to remedy the problem. 

Potential adverse events

• Potential complications related to the introduction of the catheter into the body include, but are not limited to, the following: infection, air embolism, and hematoma formation.

• Potential balloon separation following balloon rupture or abuse and the subsequent need to use a snare or other medical interventional techniques to retrieve the pieces.

• Complications associated with sizing include, but are not limited to : clot formation and embolism, nerve damage, vascular perforation requiring surgical repair, damage to the vascular intima, cerebral accident, cardiac arrhythmias, myocardial infarction, or death. 

Device Documents

Related Device SKUs

* Select 2 or more (upto 4) SKUs to compare.
SKU Name Device SKUs Model Balloon Diameter (mm) Balloon Length (cm) Introducer Size (Fr) Shaft Size (Fr) Rated Burst Pressure (atm)
PTS-X
613716 20 3 8 8 1.5
PTS-X
613719 30 3 8 8 1.0
PTS-X
613722 35 4 9 8 1.0
PTS-X
613717 20 4 8 8 1.5
PTS-X
613718 25 3 8 8 1.5
PTS-X
613720 30 4 8 8 1.0
PTS-X
613721 30 5 8 8 1.0
PTS-X
613723 35 6 9 8 1.0
PTS-X
613724 40 3 9 8 0.5
PTS-X
613725 40 5 9 8 0.5
device-thumbnail
PTS-X
  • Model : 613716
  • Procedure : -
  • Balloon Diameter (mm) : 20
  • Balloon Length (cm) : 3
  • Introducer Size (Fr) : 8
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 1.5
device-thumbnail
PTS-X
  • Model : 613719
  • Procedure : -
  • Balloon Diameter (mm) : 30
  • Balloon Length (cm) : 3
  • Introducer Size (Fr) : 8
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 1.0
device-thumbnail
PTS-X
  • Model : 613722
  • Procedure : -
  • Balloon Diameter (mm) : 35
  • Balloon Length (cm) : 4
  • Introducer Size (Fr) : 9
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 1.0
device-thumbnail
PTS-X
  • Model : 613717
  • Procedure : -
  • Balloon Diameter (mm) : 20
  • Balloon Length (cm) : 4
  • Introducer Size (Fr) : 8
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 1.5
device-thumbnail
PTS-X
  • Model : 613718
  • Procedure : -
  • Balloon Diameter (mm) : 25
  • Balloon Length (cm) : 3
  • Introducer Size (Fr) : 8
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 1.5
device-thumbnail
PTS-X
  • Model : 613720
  • Procedure : -
  • Balloon Diameter (mm) : 30
  • Balloon Length (cm) : 4
  • Introducer Size (Fr) : 8
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 1.0
device-thumbnail
PTS-X
  • Model : 613721
  • Procedure : -
  • Balloon Diameter (mm) : 30
  • Balloon Length (cm) : 5
  • Introducer Size (Fr) : 8
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 1.0
device-thumbnail
PTS-X
  • Model : 613723
  • Procedure : -
  • Balloon Diameter (mm) : 35
  • Balloon Length (cm) : 6
  • Introducer Size (Fr) : 9
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 1.0
device-thumbnail
PTS-X
  • Model : 613724
  • Procedure : -
  • Balloon Diameter (mm) : 40
  • Balloon Length (cm) : 3
  • Introducer Size (Fr) : 9
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 0.5
device-thumbnail
PTS-X
  • Model : 613725
  • Procedure : -
  • Balloon Diameter (mm) : 40
  • Balloon Length (cm) : 5
  • Introducer Size (Fr) : 9
  • Shaft Size (Fr) : 8
  • Rated Burst Pressure (atm) : 0.5